Published as a Letter on page 500 of the October 12, 1995 edition of Nature.
Authors:
R.I. Epstein, M.I. Buchwald, and B.C. Edwards of NIS-2;
T.R. Gosnell and C.E. Mungan of MST-10 at Los Alamos National Laboratory.
The possibility that an object might cool through its interaction with radiation was suggested as early as 1929 by Pringsheim [1]. After Landau [2] established the basic thermodynamic consistency of such a process, aspects of fluorescent cooling were vigorously pursued [3-11]. In particular, laser 'Doppler' cooling of gas-phase atoms and ions has today grown into a robust research area [12-15]. In contrast, attempts to cool solids with light have met with limited success; nonradiative heating effects tend to dominate, and fluorescent cooling has at best resulted in a reduction in overall heating rates [6]. Here, we report the first experimental realization of net cooling of a solid with radiation. The cooling efficiencies achieved (up to 2%) are at least 4 orders of magnitude larger than those observed in Doppler cooling of gases. By pumping a fluorescent cooling element with a high-efficiency diode laser, it should be possible to construct a compact, all-solid-state optical cryocooler; this would enable widespread deployment of cryogenic electronics and detectors in space and elsewhere [16].
The basic principles and requirements of fluorescent cooling are illustrated by the
processes occurring in our experiments which use a heavy-metal-fluoride glass
(ZBLANP) doped with trivalent ytterbium ions. Each Yb
ion possesses
only two groups of energy levels below the UV absorption edge of the host glass [17]. These
groups are separated by an energy of 1.3 eV, corresponding to a wavelength of ~1000 nm.
The ground-state group is split into four levels and the excited-state group is split into
three levels, as shown schematically in the inset of Fig. 1. The main part of this figure
shows the measured absorption and fluorescence spectra of a sample of ZBLANP doped
with 1 wt % Yb
. The mean energy of the fluorescent photons corresponds
to a wavelength 
= 995 nm, as
indicated by the vertical line. Pumping this glass in the
long-wavelength tail of the absorption spectrum moves excitations from the top of the
ground-state group to the bottom of the excited-state group. The relative populations of
the Stark levels within each group are thereby shifted slightly out of thermal equilibrium.
That is, the populations of the lower levels in each group are larger than they would be in
equilibrium. By absorbing thermal energy from the host material, thermal
equilibrium within each group is restored. Radiative decays from the excited- to the
ground-state groups produce photons that carry off the absorbed radiative and thermal
energies.
Figure 1:
Main plotAbsorption coefficient
(dashed curve) and fluorescence spectrum (solid curve) at room temperature of
ZBLANP
(ZrF
-BaF
-LaF
-AlF
-NaF-PbF
)
doped with 1 wt %
Yb
. The actual
peak absorption coefficient is 1.1 cm
at
975 nm. The fluorescence spectrum was recorded using a one-meter spectrometer
and an InGaAs detector in a 90 degree scattering geometry, and was corrected for the nonuniform responsivity
of the system. The shape of the fluorescence spectrum is the same regardless of where
one pumps in the
Yb
absorption band.
The wavelength corresponding to the mean fluorescent-photon energy, 

,
is indicated by a vertical line. Inset plotThe schematic energy-level structure of
ZBLANP:Yb
; the splittings within
each group have been exaggerated for clarity. The arrows denote a typical cooling cycle: (1) laser pumping,
(2) energy absorption from the thermal bath, (3) radiative decay, and (4) additional thermal-energy absorption.
Successful fluorescent cooling requires that a negligible fraction of the decays from the excited- to the ground-state groups occurs through nonradiative, heat-generating processes. Cooling is, at best, proportional to the energy spread of each group, whereas nonradiative decays generate heat in proportion to the much greater energy spacing between the groups. A high radiative quantum efficiency is thus required in order that nonradiative heating not overwhelm the fluorescent cooling. Kastler [3] suggested that rare-earth ions in transparent solids could be effective fluorescent coolers because of their large quantum efficiencies.
Each fluorescent photon carries off, on average, thermal energy equal to the difference
between the pump-photon and the mean fluorescent-photon energies. In the ideal case
where there are no nonradiative relaxations from the excited- to the ground-state groups,
the cooling power, P
, is proportional to the absorbed
pump power, P
, and to the average difference in the photon
energies of the pump and fluorescence radiation. In
terms of wavelength
of the pump radiation, the cooling power is thus

.In our work, two different experimental arrangements are used to investigate the
fluorescent cooling of ZBLANP:Yb
. In the first of these,
photothermal-deflection spectroscopy [18-19] is employed to measure the local
temperature gradients induced by the pump laser in the interior of a sample. The pump beam from a c.w.
titanium-sapphire laser is focused into the sample. A helium-neon laser probe beam, coaligned with and
slightly displaced from the pump beam, passes through the sample in the opposite
direction. Angular deflections of this probe beam, which are caused by thermally-induced refractive-index
gradients in the pumped volume of the sample, are measured
with a position-sensitive photodetector. An optical chopper placed in the path of the
pump beam modulates the photothermal-deflection signal.
Figure 2: (a) Photothermal-deflection waveforms recorded on an averaging oscilloscope for pump-laser wavelengths of 980 and 1010 nm, at a chopping period of 1.8 s. This period is long compared to the relevant atomic and thermal relaxation time scales of the sample. The corresponding laser powers are 1.03 and 0.70 W, respectively, with a focused beam diameter of ~50 mm. The sample length is 2.7 cm. The ~50 ms rise and fall times depend upon the lateral separation between the pump and probe beams. (b) Amplitudes (filled circles) of the photothermal-deflection waveforms normalized by the incident laser power as a function of pump-laser wavelength. The solid curve is a plot of Eq. (1), scaled by an arbitrary temperature-to-deflection conversion factor.
Two examples of raw photothermal-deflection data recorded on an averaging
oscilloscope are shown in Fig. 2(a). This panel presents deflection waveforms
synchronous with the chopper for pump wavelengths of 980 and 1010 nm, respectively
15 nm below and above 
. The
unmistakable 180 degree phase difference between the two
waveforms indicates that the positive temperature gradient observed at 980 nm becomes a
negative temperature gradient at 1010 nm. Evidently, the probed volume of the sample is
cooling at the longer wavelength.
A more systematic investigation of the photothermal-deflection spectra appears in Fig. 2(b).
In this figure, the filled circles specify the measured deflection amplitudes normalized
with respect to the incident pump-laser power. A clear transition from the heating to the
cooling regime is observed as the pump wavelength is tuned to values larger than

,
indicated by the vertical line.
In the second experimental arrangement, the equilibrium temperatures reached by a bulk
2.5 mm by 2.5 mm by 6.9 mm sample are measured when the sample is continuously
pumped by a titanium-sapphire laser beam. The sample is positioned at the center of a
vacuum chamber and rests on two thin, vertical glass slides that thermally insulate the
sample from the chamber. The inner wall of the chamber is painted black to absorb stray
pump radiation as well as the fluorescence emitted by the sample. The pump-laser beam
is directed along the long axis of the sample. The dimensions of the sample transverse to
the optical axis are small enough that self-absorption of the fluorescence is minimal. The
temperature of the sample is monitored with a liquid-nitrogen-cooled InSb infrared
camera which is sensitive in the 3-5 µm range but blind to the ~1 µm laser and
fluorescence radiation. Because ZBLANP is transparent at wavelengths shorter than 5
µm, a 1 mm
square of gold foil painted black on its outer face is
attached to the
sample to enhance the emissivity at the detection wavelengths. The outer surface of the
foil emits thermal radiation characteristic of the temperature of the sample, while the
polished inner surface reflects most of the fluorescent radiation. To account for
the changes in the sample temperature arising from temperature drifts of the chamber, a
reference sample with an identical gold foil is positioned ~1 cm away from the test
sample, outside the pump-beam path but on the same glass supports in the chamber.
Temperature differences of 0.02 K between the pumped and reference samples can be
resolved. When the test sample is exposed to the pump laser, we find that the sample
temperature equilibrates in about 15 min.
Figure 3: Equilibrated temperature differences (filled circles) of a laser-pumped 0.69-cm-long bulk sample relative to an identical reference sample. Measurements are made with an infrared camera thermometrically calibrated by comparison to a type-T thermocouple. Temperature changes are normalized by the incident laser power. The values between 1008 and 1030 nm are less than zero, corresponding to cooling. The solid curve is a plot of the expected temperature changes from Eq. (2), reduced by 20% to allow for laser radiation which misses the sample. The differences in the data near the peak at 975 nm, compared to those of Fig. 2(b), are due to differences in sample lengths and hence optical depths.
The filled circles in Fig. 3 show the steady-state temperatures of the test sample relative to the reference sample as a function of the pump wavelength per watt of incident laser power. Note that negative temperature differences as large as 0.3 K are obtained at long pump wavelengths. These results definitively demonstrate, for the first time, the absolute cooling of a solid by laser-induced anti-Stokes fluorescence.
The equilibrium temperature, T
, of the sample is established by
a balance between the laser-induced fluorescent cooling and the heat load from the environment. In our setup,
the dominant heat load is from the radiative coupling between the walls of the vacuum
chamber at ambient temperature T
and the sample. If the
sample radiates as a blackbody, then

,where
is the Stefan-Boltzmann constant and A = 0.82
cm
is the surface area of the
sample. For small temperature changes, Eq. (2) implies
,with P
given by Eq. (1).
This prediction is plotted as the solid curve in Fig. 3, scaled vertically by 0.8.
The small discrepancies between theory and experiment in the cooling region can be reconciled if the radiative
quantum efficiency equals ~0.997 and if ~0.1% of the laser power directly heats the sample. The results of
the two types of experiments are in good agreement with each other, as can be seen in
Fig. 4, where the data from Figs. 2(b) and 3 have
been normalized by the corresponding
absorptivities and plotted together for comparison. This normalization gives the cooling
efficiency, defined as the ratio of the cooling power to the absorbed laser power. In
agreement with Eq. (1), most of the data lie on a straight line. If the quantum efficiency
were unity, the zero crossing would occur at 
= 995 nm.
Experimentally, the zero crossing agrees with this prediction to within 3 nm, indicating that the radiative quantum
efficiency is at least 0.997. The deviations from linearity for
1020 nm may be
explainable in terms of parasitic heating in our sample or experimental artifacts.
The experimental cooling efficiency in Fig. 4 is ~2% at a pump wavelength of 1015 nm.
Figure 4: Cooling efficiencies measured in the
photothermal-deflection (filled circles) and bulk-cooling (open squares) experiments. Negative efficiencies
correspond to heating. The amplitudes of the two data curves have been adjusted using the scaling factors from
Figs. 2(b) and 3. The solid curve is a plot of
P
/ P
from Eq. (1).
A computer model of the excitations and radiative transfer in the ZBLANP:Yb
system predicts that similar efficiencies should be achievable down to a temperature of 60 K.
By pumping a fluorescent cooling material with an efficient diode laser, an all-solid-state cryocooler
could be developed that operates with an efficiency comparable to commercial mechanical coolers [16].
This type of cryocooler would be functional in the temperature range useful for high-T
superconductors, infrared detectors, and other cooled electronic devices and would be well-suited for space-based applications.
[2] Landau, L., On the thermodynamics of photoluminescence, J. Phys. (Moscow) 10, 503-506 (1946).
[15] Chu, S., Laser manipulation of atoms and particles, Science 253, 861-866 (1991).
[16] Edwards, B.C., Buchwald, M.I., Epstein, R.I., Gosnell, T.R., & Mungan, C.E., Development of a fluorescent cryocooler, Proceedings of the Ninth Annual American Institute of Astronautics & Aeronautics Utah State University Conference on Small Satellites (ed. Redd, F.) in press (Utah State University, Logan, 1995).
[17] Dieke, G.H., Spectra and Energy Levels of Rare Earth Ions in Crystals (Interscience, NY, 1968).
We thank D.E. Casperson, C. Edwards, C. Ho, N. Kurnit, S. Lloyd, A.V. Olinto, W.C. Priedhorsky, W.R. Scarlett, and P. Xie for helpful discussions and comments. A.J. Gibbs and J. Fajardo are appreciated for help in preparing the glass samples. This work was carried out under the auspices of the U.S. Department of Energy.